Kibion® Dynamic
Do I have to regularly check the Kibion® Dynamic instrument?
No, there is no user required maintenance necessary. The Kibion® Dynamic should be checked once per year by a Kibion service technician or a qualified service technician of a dedicated Kibion distributor. For flawless function Kibion recommends a yearly service including change of filter.
How often does the Kibion® Dynamic base need to be auto-adjusted?
The daily auto-adjustment has to be performed every day. The weekly auto-adjustment has to be performed once a week.
When I try to perform a weekly auto-adjustment and I get a “Low CO2” error. How can I fix it?
The weekly auto-adjustment requires a breath sample with at least 3,5 Vol. % CO2 to cover the full range of samples. If you experience problems, make sure that:
- You hold your breath for 30 seconds or longer
- You use a bag with a volume >1 L
If you still encounter this issue try with another user or contact Kibion GmbH Technical Service or your local Kibion GmbH representative.
I receive an “No connection to the internal IO-Board possible” and/or “No connection to the analyzer possible” when I start a measurement. What can I do?
Turn off the instrument and unplug the power cable for 5 minutes. Turn it on again and wait for 10 minutes before you log in.
If you still receive the error please contact Kibion GmbH Technical Service or your local Kibion GmbH representative.
Disposal of waste electrical equipment
Please follow your local regulation and good laboratory practices.
Heliprobe® System
What to do if the BreathcardTM is defective?
If the BreathcardTM is defective (e.g., damaged lithium patch, indicator already yellow, mylar foil torn, etc.), please dispose of it according to good laboratory practices and local regulations and use a new one.
What to do if the Heliprobe® Analyzer is defective?
Please contact Kibion GmbH Technical Service or your local Kibion GmbH representative.